The gloveboxhas beautiful design, compact structure and fine workmanship
Theinert gasin the glove box is circulated in a closed cycle by a circulation fan and a purifier, continuously absorbing O2 and H2O
TheH2O and O2removal materials can be regenerated, and the regeneration process is automatically controlled by the program
Using thelatestEuropean technology, all key components are imported from Europe, from world renowned manufacturers
The controlsystemincludes self-diagnosis, power-off and self starting features, pressure control and adaptive functions, password protection, and the control unit uses Siemens PLC touch screen
The water analyzer has strong corrosionresistanceand wide application fields, especially for users of lithium battery manufacturing and metal organic, etc. it can be cleaned for 3 times and reused repeatedly, avoiding the problem of scrapping once pollution
The oxygen analyzeruses the ZrO2 sensor to avoid the problem of short life of the fuel cell and continuous exposure to the air
The use of integrated valve structure reduces and optimizes the system pipeline layout, reduces leakage points, simplifies the leak detection link, and the use of products is more stable and reliable
The equipmentruns stablyand can run continuously for 7 × 24h
Automatic control, simple operation, low operation and maintenance cost
Specialized, standardized and large-scale production of all parts of the glove box
Provide products for customers with various special requirements
All parts have been strictly audited and controlled
Rich design and control experience
Theproduction process is strictly based on the ISO9001 standard
Lab2000 2400 Split Double-sided Glove Box for Anaerobic Glove Box
Review the listed configuration, connection interfaces, installation space, operating workflow and maintenance access for this product. Confirm the final scope against the actual project before selection.
TechnicalParameters
Chamber
Description
Material: Stainless steel 1.4301 (SUS type 304); Thickness 3 mm;Dimensions: 2 4 0 0 mm(94.5″) x 750mm(29.5″)x 900mm(35.4″) L x W x H
Volumn
0.8 m³(28.2cu)
Front window
Panel: 8 mm thick safety tempered glass or 10 mm thick polycarbonate (optional);Dimensions: 1123mm (44.2″) x 839mm (33.0″) L x H
Glove port
Hard aluminum alloy material or polyoxymethylene material (optional); Glove port diameter: 220 mm (8.7″); O-ring seal
Gloves
Material: Butyl rubber; Thickness: 0.4 mm or 0.8 mm (available for selection)
Filter
Outlet and inlet filters, filter rate<0.3μm
Illumination
LED lighting is located on the top of the front window
Leak rate
Final acceptance test according to the standard of leakage rate <0.05 vol% / h
GP20 Gas Purification System
Description
Automatic removal of H2O and O2; single purification column system, automatic regeneration (optional GP200 double purification column); closed gas circulation line
Use voltage
AC 230 V / 50-60 Hz, 10 A or AC 115 V / 50-60 Hz, 20 A (Available for selection)
Working gas
Working gas:N2、Ar 、He(purity≥99.999%);Regeneration gas:H2 accounts for 5% -10%, the rest is working gas (purity ≥99.999%)
Vacuum pump
Specifications: Rotary vane vacuum pump, equipped with oil mist filter, with gas ballast control; flow rate: 12 m³ /h (7 cfm), vacuum degree < 2 x 10-3 mbar (dry pump optional)
Circulation unit
Oil-free high-speed fan; air volume: 0-100 m³ / h (0-59 cfm)
Valve
Electromagnetic high vacuum valve
Leak rate
Final acceptance test according to the standard of leakage rate <0.05 vol% / h
Cleaning system
Description
After setting the corresponding time and pressure, the system can automatically perform gas replacement in the chamber.
Analyzer
Oxygen analyzer
Size
205 mm(8.1″) x 80 mm(3.1″) x 60 mm(2.4″)L x W x H
Measurement range
0-1000 ppm
Other oxygen transmitter
GE oxy.IQ™ Oxygen Transmitter
Dew analyzer
Size
205 mm(8.1″) x 80 mm(3.1″) x 60 mm(2.4″)L x W x H
Measurement range
0-500 ppm
Other watertransmitter
GE VeriDri™ Dew-Point Transmitter
Solvent purification system
Leak rate
Material: Stainless steel 1.4301 (SUS type 304), thickness 3 mmInternal size: 220mm (8.7″) x 450mm (17.7″) Φ x HHigh-quality activated carbon
Optional equipment
Refrigerator
Independent control system, integrated in the side panel of the chamber, temperature -35 ℃, volume 18L or 32L optional; main chamber 304 stainless steel sheet, 5 layers of variable compartments R404 environmentally friendly refrigerant, imported
Heating chamber
The main antechamber is equipped with a heating system with a temperature of 200 ℃ and a temperature control of ± 1 ℃
Other instructions
Productcertification
ISO9001 certification, CE certification, UL certification
Warranty
One-year warranty period, lifetime maintenance
Applicationconsiderations
For details, please refer to the instruction manual danger, warning, caution, etc